GET THE APP

Electrochemical sensing of cytochrome c using Graphene Oxide | 80105

Jornal Internacional de Pesquisa Inovativa em Ciência, Engenharia e Tecnologia

Abstrato

Electrochemical sensing of cytochrome c using Graphene Oxide nanoparticles as platform

Bhawna Batra , Samiksha Sangwan , Jyoti Ahlawat , Minakshi Sharma

An improved cytochrome c (Cyt c) biosensor based on immobilization of cytochrome c oxidase (COx) on the surface of graphene oxide nanoparticles (GONPs) electrodeposited onto pencil graphite (PG) electrode. Characterization of graphene oxide nanoparticle was done by Transmission electron microscopy (TEM), Fourier transform infra-red spectroscopy (FTIR) and X-ray diffraction study (XRD). The working electrode (COx/GONPs/PG) was characterized at its different stages of fabrication by scanning electron microscopy (SEM) and FTIR. Fabrication of Cyt c biosensor was done by connecting COx/GONPs/PG as working electrode, Ag/AgCl as reference electrode and Pt as auxiliary electrode to potentiostat. The mechanism of detection of present biosensor was based on oxidation of Cyt c (reduced) to Cyt c (oxidized) by COx resulting in flow of electrons through GONPs to the PG electrode, hence current generated is proportional to the concentration of Cyt c. Present biosensor exhibited optimum potential at 0.49 V with optimum pH 7.5 and optimum temperature 35°C. Biosensor showed linearity within 40–180 ng/ml having 40 ng/ml limit of detection.

Isenção de responsabilidade : Este resumo foi traduzido utilizando ferramentas de inteligência artificial e ainda não foi revisado ou verificado.